Technical Specifications |
VERTICAL DEFLECTION (Y) |
Deflection Coefficient (CH1 & CH2) |
1mV/div to 10V/div. 5mV/div to 10V/div in 5 calibrated steps
in 1-2-5 sequence. x5 Magnification increases the sensitivity to
1mV/div to 1V/div. |
Accuracy |
x1 : ±3%, x5 : ±5%. |
Variable |
1 / 2.5 times uncalibrated continuously variable control
extends the Deflection Coefficient to more than
12.5V/div. |
Bandwidth |
x1 : DC to 60MHz (-3dB), dc coupled. 10Hz to 60MHz
(-3dB), ac coupled. x5 : DC to 7MHz (dc coupled) 10Hz to 7MHz
(ac coupled). |
Rise Time |
6 ns or less. |
Display Modes |
CH1 only, CH2 only, CH1 & CH2 alternate or chop mode,
Algebraic additon CH1+CH2, Algebraic subtraction CH1 - CH2, CH2
invert & X-Y. |
Input Impedance |
1M ohms & 25pF approx. |
Maximum Input Voltage |
300 Volts (dc + peak ac). |
Internal Trigger Signal |
CH1 or CH2. |
TIME BASE |
Sweep Speed |
20 Calibrated steps. 0.1µs/div to 0.2s/div in 1, 2 & 5
sequence. |
Sweep Magnifier |
x10 Magnification increases sweep speed to
10ns/div. |
Accuracy |
±3%. |
Variable |
Uncalibrated continuously variable control between steps,
extends slowest sweep speed to 0.5s/div (approx.). |
ALT MAG Trace |
x10 magnified sweep is displayed along with normal
sweep. |
TRIGGER SYSTEM |
Triggering Mode |
AUTO, NORM, TV-V, TV-H. |
Source |
CH1 / CH2 / LINE / EXT. |
Slope |
Positive or Negative. |
Coupling |
ac coupling. |
Trigger Sensitivity |
|
Internal |
Auto : 2 div - 20Hz to 60MHz. Normal : 2 div - 10Hz to
60MHz. Lock : 3 div - 50Hz to 30MHz. |
External |
Auto 0.2V p-p - 20Hz to 60MHz. Normal 0.2V p-p - 10Hz to
60MHz. Lock 0.3V - 50Hz to 30MHz. |
HORIZONTAL DEFLECTION |
Deflection Coefficient |
Same as CH1. |
Bandwidth |
DC - 500KHz (-3dB). |
Input Impedance |
1M ohms & 25pF approx. |
Phase Difference |
<3º (DC - 50KHz). |
GENERAL INFORMATION |
Cathode Ray Tube |
140mm Rectangular screen, Internal Graticule, 8 x 10 cm, P31
phosphor. Accelerating potential : 12KV. |
Z-Modulation |
±5V p-p signal upto 2MHz modulates at normal
intensity. |
Calibrator |
Provides 0.5V ±2%, 1KHz ±2% square-wave output for probe
compensation. |
Power Requirement |
220V AC ±10%, 50Hz ±5%, 40VA (max.). |
Dimensions |
140 (H) x 335 (W) x 375 (D) mm approx. |
Weight |
7.2 Kgs. approx. |
Accessories |
Instruction Manual, 2 High impedance switch probe with x1 or
x10 attenuation, Mains Cord. |